Calibrate a pH meter with bracketing buffers
Practice the USP <791> pH-meter order. Samples are not read until a two-or-more-point calibration brackets the expected pH and an independent check buffer passes.
- Inspect the electrode.
- Confirm the fill solution is at the marked level.
- Calibrate with fresh buffers that bracket the expected sample pH.
- Rinse between buffers.
- Allow a stable reading before accepting each point.
- Verify with an independent check buffer.
- Measure the sample only after the check is within tolerance.
- Do not use a one-point “tweak” in place of a failed calibration.