Autoclave a biohazard load
Practice the sequence you already learned for “Autoclave a biohazard load” until the order feels automatic. Drill the BMBL / AAMI steam-sterilization order. Bags stay loosely closed so steam can penetrate, and the load is not signed off until the indicator has passed.
- Confirm the load type.
- Select the matching validated cycle.
- Load so steam can penetrate; do not overpack sealed bags.
- Place a chemical integrator in the densest part of the load.
- Run the validated cycle; do not interrupt it.
- Wait for the chamber to depressurize before opening.
- Unload with heat-resistant PPE.
- Read the integrator or biological indicator.
- Quarantine the load if the indicator fails.
- Log the cycle, operator, and result.